03 Series Operator’s Manual
Calibration Mode • 44
adjustment (zero adjustment for OX-03).
10. If the span adjustment (zero adjustment for OX-03) is successful, the
LCD will show the following screen before returning to the E--CAL
screen.
If the Maximum Span setting is turned On (factory setting is Off) using
the 03 Series User Setup Program, the LCD will show the maximum
reading the sensor could have been calibrated to before returning to the
E--CAL screen.
11. If the span adjustment (zero adjustment for OX-03) fails, the LCD will
show the following screen, the alarm LED will flash, and the buzzer will
sound a double pulsing tone.
Press and release the POWER MODE button to clear the failure
indication and return to the E--CAL. screen.
See “Troubleshooting” on page 63 to investigate the cause of the failure
and replace the failed sensor if necessary.
12. Turn the regulator on/off knob clockwise to close it.
13. Remove the calibration adapter from the instrument.
PASS
C--MAX
ppm
75
FAIL
CO
ppm
E--CAL.