Operating Manual Version 7.3-03 of 28.07.20
8
Load L1, S2 – ‘On’,
load capacity – 66 nF
Fig.4
Load L2,
load capacity – 2.2 nF
Fig.5
2.3.3 Amplitude modulation (see Fig.6) with the following settings:
pause time: (1.0 ± 0.5) sec;
stimulus burst time:
in the 1:1 mode – (1.0 ± 0.5) sec;
in the 2:1 mode – (2.0 ± 0.5) sec;
in the 3:1 mode – (3.0 ± 0.5) sec;
in the 4:1 mode – (4.0 ± 0.5) sec;
in the 5:1 mode – (5.0 ± 0.5) sec.
U
min
– minimum amplitude
U
s
– set amplitude
t
p
– pause time
t
b
– stimulus burst time
Fig.6
2.3.4 BEE mode
– generation of a single stimulus with the first
pulse of the 2
nd
phase having the highest amplitude;