Operating Manual Version 7.3-03 of 28.07.20
24
6.2.4 Setting the Damping mode (
Dmp
)
The Damping modes are the following:
Damping Off
(
Dmp Off
)
is indicated by a long beep;
Damping 1
(
Dmp Sс1
)
is indicated by a short beep;
Damping 2
(
Dmp Sс2
)
is indicated by 2 short beeps;
Damping 3
(
Dmp Sс3
)
is indicated by 3 short beeps;
Damping 4
(
Dmp Sс4
)
is indicated by 4 short beeps;
Variable damping
(
Dmp Var
)
– is indicated by a two-tone beep.
6.2.5 Setting the Frequency (
F
)
6.2.5.1 To control the frequency, you can either press the buttons
‘step by step’ or press and hold the appropriate button (this way is faster).
Frequency can be controlled within the range 15 (
15.6
) to
350 (
354
) Hz step by step at a step of about 2 % of the current frequency
value (128 steps in all). When the maximum or minimum frequency is
reached, the device emits a long beep.
In addition, reaching the frequencies of
30.5
,
59.6
,
90.7
,
119
and
181
Hz is indicated by 1, 2, 3, 4, and 5 short beeps respectively. While
the beeps sound, the frequency cannot be changed. This makes it easier to
set the required frequency.
6.2.6 Setting the Intensity (
Int
)
The intensity (the number of stimuli in a burst) can be controlled with-
in the range of 1 to 8 but the maximum intensity value depends on frequen-
cy, gap, and energy strength. The greater the frequency, gap and energy
strength are, the smaller maximum intensity you can set. For example, at a
maximum frequency (
354
Hz) the intensity cannot be greater than 2.
With increased frequency, gap or energy strength, the intensity is au-
tomatically decreased when needed.
When the intensity is more than 1 apart from the main frequency
(which can be set in the item
‘F’
of the Main Menu), an additional
(equivalent) frequency appears, which is higher than the set one and de-
pends on the stimuli energy and Gap, see Table 2.
The intensity of the additional (equivalent) frequency in relation to
the main directly depends on the item
Intensity
of the Main Menu.
(Optional) The device can indicate the equivalent frequency during