Operating Manual Version 7.3-03 of 28.07.20
14
5 PREPARING DEVICE FOR OPERATION
5.1 Before every procedure disinfect electrodes with a cotton wad
wetted in a 3 % hydrogen peroxide solution with the addition of 0.5 %
solution of an approved cleaning liquid. Allow to dry up thoroughly be-
fore use. Clean case of device and add-on electrodes with a damp (not
soaking) cloth and mild soap solution. Allow to dry before use.
5.2 Open the battery compartment cover and
install the batteries. Switch the device on using
the switch button on the lateral panel. A message
containing the information about the device’s
name, version and its software release date
(Fig.15) will be displayed for 2 seconds and a
beep will sound.
5.3 Then the device switches to the state that
displays
basic parameters
(hereinafter called
‘B’
state
). So, the basic parameters of stimulation
will be displayed (Fig.16).
Fig.15
The first line contains the timer, indica-
tor of the device’s contact with the patient’s
skin (
●
– contact,
○
– no contact), and a
battery charge indicator.
The timer is reset
upon pressing any button, and when au-
tomatic turn-off is enabled, the timer is
reset on contacting the skin as well.
5.4 Press the
button to switch to
the
Menu state
(Fig.17). Press the
button at an interval of no longer than
2 seconds to make sure that the highlight-
ing switches cyclically.
Then wait until the device switches to
the
‘B’ state
(this will happen in
2 seconds after the latest pressing of the
button
), then press the
–
button once.
A long (1 sec) beep will sound.
Fig.16
Fig.17
Press and hold the button
+
. The number in the right part of the top
line will be changing from 1 to 250, and then a long beep will sound.