Operating Manual Version 7.3-03 of 28.07.20
17
Attention!
To avoid painful sensations
when treating the most sensi-
tive areas on the patient’s body, it is recommended that the
energy strength is decreased to minimum before the treat-
ment by pressing and holding the
–
button until a long
beep sounds).
6.1.6 Apply the electrode on the skin on the zone to be treated. The
●
symbol
will be displayed on the screen indicating that the electrode is
contacting the skin. Make sure that patient isn’t experiencing unpleasant
sensations. Press and hold down the
+
button till the first sensation like
pricking, light burning or vibrations appear. They shall be sensed as com-
fortable and not painful, but must be felt on the skin.
Attention!
When using add-on electrodes, connect electrode cable
only to the switched-off device, in order to prevent the de-
vice damage and to avoid painful sensations!
6.1.7 To start therapy, set the required mode. Then put the electrode
on the skin area to be treated (see the Instruction for Use provided with
your device).
6.1.8 To treat auricular zones and acupuncture points, connect your
add-on point electrode to the appropriate jack. Make sure of the contact
by touching the skin with the electrode. Before every treatment of any
acupuncture point, dip the end of the probe in water to wet the surface.
Stimulation by point electrode that is too small or by any electrodes
that applied incorrectly could result in discomfort.
Note
When using add-on electrodes, especially point and comb, the
dosing mode can be out-of-work even in case of good contact with
skin and even if the patient feels the stimulation.
It is not a fault.
In the
‘B’
state
the
+
and
–
buttons are used to control the energy
strength within the range from 1 to 250 units. To control the energy
strength you can either press the buttons ‘step by step’ or press and hold
the appropriate button (this way is faster, 16 steps per second). When the
energy strength reaches the value that cannot be further in-
creased/decreased, the device will indicate it with a long beep.
6.1.9 If the automatic turn-off is enabled (see p. 6.3.1), and if there is
no contact with the skin for 30 seconds, the device will automatically
switch to the standby mode. This will be accompanied by a beep, text
message and screen darkening. To resume SCENAR operation, press the
button.