Once the device detects that the electrode is contacting the skin, the
device will give a short low pitch signal (
○
will change to
●
in the top
line), and in a second it will give a short high pitch signal and the screen
will display (both digitally and graphically) the results measured over the
first second (the diagram will show current (ongoing) reaction change
with time).
After 1 second more, measurement results for the 2
nd
second will be
displayed on the screen. Then,
current
results of measuring (time, reac-
tion, rate of reaction change, shape factor) illustrated in the Fig.21 will be
updated every second, and the diagram will be filled up. It will continue
this way until either
Dose
or
Zero
is achieved.
Timer
Initial reaction
Initial shape factor
‘Dose’
Current shape factor
Reaction rate
Reaction and shape
rate chart
‘Zero’
‘Dose’
‘Zero’
Current reaction
Fig.21
In
Dose 1
,
until
Dose
is delivered, the device emits beeps indicating
the time elapsed from the moment the electrode contacted the skin. At the
10
th
second the device emits one short beep, at the 17
th
sec – two beeps, at
the 24
th
sec – three beeps, and so on – n beeps at the (3+7*n) second.
When the
Dose
is reached it is simultaneously indicated by:
2-second dual-tone beep and;
the
*
symbol displayed along with the values of the reaction and
shape factor registered at a point of the Dose (in red).
When the
Zero
is reached it is simultaneously indicated by:
series of low pitch sounds during 2 seconds;
the
○
symbol displayed along with the values of the reaction and
shape factor registered at a point of the Zero (in blue).