Measuring Ring Amplitude
Now let’s measure the amplitude of the ringing. To measure the amplitude, do these
steps:
1. Press CURSOR
key to see the Cursor menu.
2. Press Mode to set
Manual
mode
3. Press Type to select
Voltage
.
4. Rotate vertical POSITION
knob to place cursor A on the top of the first positive
peak of the ring.
5. Rotate horizontal POSITION
knob to place cursor B on the bottom of the first
negative peak of the ring.
You can see the following measurements in the cursor menu: (See Figure 3-5)
■
The delta voltage (peak-to-peak voltage of the ringing)
■
The voltage at Cursor 1
■
The voltage at Cursor 2
Figure 3-5