Understand the vertical system
Ⅰ
.Settings of the channels
Each channel of DS5000 has an operation menu and it will pop up after pressing
CH1 or CH2 button. The settings of all items in the menu are shown in the table
below.
Table2-1
Menu
Settings
Comments
Coupling
AC
DC
GND
AC blocks the DC component of the input Signal
DC passes both AC and DC components of the input
signal
GND disconnects the input signal
BW Limit
ON
OFF
Limits the channel bandwidth to 20MHz to reduce
display noise.
To set “OFF” will get full bandwidth.
Probe
1X
10X
100X
1000X
Set this to match your probe attenuation factor to
make the vertical scale readout correct
Digital filter
Setup digital filter
Volts/Div
Coarse
Fine
Selects the resolution of the SCALE knob coarse
defines a 1-2-5 sequence.
Fine changes the resolution by small steps between
the coarse settings.
Invert
ON
OFF
Turn on the invert function.
Restore to original display of the waveform.
Input
1M
Ω
50
Ω
Set 1M
Ω
input impedance
Set 50
Ω
input impedance
NOTE:
The input impedance select function (1M
Ω
/50
Ω
) is only provided in certain
models.