To use average follow these steps.
Copyright 2004 Rigol Electronic
3-5
·
Press soft button as ACQUIRE
→
Acquisition
→
Average
·
Toggle the Averages soft button to select the number of averages that best
eliminates the noise from the displayed waveform. The higher the number of
averages is, the more noise that is removed from the display. However, the
higher number of averages, the slower the displayed waveform responds to
waveform changes. You need to choose between how quickly the waveform
responds to changes and how much noise there is on the signal. (See Figure
3-3)
Figure 3-3