66 -
|
Page Rigel Medical
62353+
user Manual V1.4
6 Trouble Shooting
To ensure the measurements are performed under valid and consistent conditions, the
the Rigel 62353 Plus is programmed with the ability to perform a number of self checks
and test setup verification checks upon power up and prior to a powered safety test
(leakage test).
The Rigel 62353 Plus will test the incoming mains supply as follows:
L-N polarity of incoming mains (UK and polarised mains versions only)
o
Test may continue
Integrity of the incoming earth (TN supply) (see application note 0002)
o
Test may continue
Check for over voltage on Live and Neutral wire
o
Test cannot continue
Example of IT warning (incoming earth fault);
In addition, the Rigel 62353 Plus is able to check for a valid test setup for
measurements done using the direct method. In case a secondary earth path is
detected, the Rigel 62353 Plus will warn the user. It is
not possible to continue
as
testing under such conditions will always result in a “zero” reading due to the low Ohm
resistance of the secondary earth path in comparison to the 1000 Ohm resistance of
the body model used in the Rigel 62353 Plus.
Example of Secondary Earth Path warning: