31 -
|
Page Rigel Medical
62353+
user Manual V1.4
3.5 Trace Variables
To increase the traceability of the safety test, the user can include valuable data with
the test record such as Re-test period and Asset variables (see 2.2 and 3.1 for more
information)
In this example, all Asset Variables are activated.
8. Use the Up & Down Keys to navigate the menu and the Left & Right Keys to
open the drop down boxes [<>]. Select the required input and confirm with the
button (F4). Each variable entered during the automatic test will
automatically be added to the list for future use. Trace Variables that are de-
activated (set to NO in the Setup menu, section 2.2) will not appear during the
automatic tests.
A maximum of 40 individual entries are available for each Trace
Variable. Additional entries will overwrite on a first in first out basis.
When all information is provided, press the
button (F4) to start the safety test.
To ensure the safety of the user, non-electrical tests are done prior to the powered
electrical safety tests.
Non-powered tests are deemed;
Visual and Custom tests (if applicable),
Earthbond test (class I)
Insulation test (if applicable)
In this example we will perform a safety test on Class I equipment, using a preset test
sequence, based on IEC 62353 Plus with added Visual Test prior to the safety test and
a Custom Test followin the successful safety test.
The Applied Part Configuration is set to default and the test is executed in Semi
Automatic mode.