16 -
|
Page Rigel Medical
62353+
user Manual V1.4
Once the Insert button (F1) has been pressed, a drop down box will appear with all
available safety and inspection tests. Activate the drop down box using the left key and
or scroll through the available tests using the up and down keys.
Once the desired test has been highlighted, confirm using the OK button (F4) and finally
confirm to insert into the test Sequence. Pressing the
button (F3) will return to
the previous screen without changes being made.
The individual test has now been inserted and can be edited as described in 2.1.4.
Further electrical safety tests can be inserted or deleted by following the steps
described herein.
Proceed with non-electrical safety tests as described in the following or save the new
Test Sequence by following the steps below.
2.1.6 Insert a non-electrical safety test
Non-electrical safety tests are intended to allow the user to capture additional
information either prior to or after a safety test is completed. Such information could