Do not exceed the maximum permitted voltage of 30 V AC/DC
with respect to earth potential! Electric Shock danger!
Press
button (F3) to set the Test Duration, and RMS Pass/Fail limit. Use the
arrow keys to highlight and set. Press the
button (F4) when complete. Pressing
the
button (F3) will exit Settings mode without saving changes.
Press the
button (F2) to change the mains polarity.
For both Class I and Class II appliances, connect ALL* Patient Connections or Applied
Parts to the Applied Part Module and fit the Appliance mains plug into the EUT outlet.
All Patient Connections and Applied Parts are grouped together which
eliminates the requirement to configure the Applied Part Module as
the Rigel 62353 Plus will short together all 10 connections as part of
this test. For this reason, the Applied Part configuration key is not
available.
For Class I equipment use the earth bond lead (from the black socket) and connect to a
conductive part of the enclosure. For measurements on conductive non-earthed parts,
the test needs to be repeated using the same probe.
For Class II devices use the earth bond lead (from the black socket) and connect to the
enclosure preferably encapsulated within conductive foil.
Press the green START key to begin the test. The test will run for the duration set or
until the red STOP key is pressed.
4.1.6 Equipment Leakage (Differential)
The Equipment Leakage Test measures the total leakage deriving from the Applied
Parts, Enclosure and Mains Parts combined to real earth.
This test is applicable to Class I and Class II B, BF and CF equipment. This is an IEC
62353 Plus test. For a complete description of this test and applicable pass / fail limits,
please see Appendix B.