4.1.8 Applied Part Leakage (Direct)
The Applied Part Leakage Test measures the total leakage deriving from the combined
Patient Connections within an Applied Part to earth and any conductive or non
conductive parts on the enclosure (either connected or isolated from earth) under the
fault condition Mains on Applied Parts.
The Applied Part Leakage Test is applicable to class I and class II floating type (BF &
CF) Applied Parts. This is an IEC 62353 test. For a complete description of this test and
applicable pass / fail limits, please see Appendix B.
Press the
button (F1) to obtain the schematic representation of this test:
Warning
– never leave the earth bond, or any other probe attached to any
moving parts.
Warning
– never touch the appliance under test whilst testing is in
progress. Doing so could result in injury due to electric shock and / or
sudden movement of any mechanical parts on the appliance.
Refer to Page 4 for notes on precautions required for compliance with the
relevant safety standards.
Do not exceed the maximum permitted voltage of 30 V AC/DC
with respect to earth potential! Electric Shock danger!
Press the
button (F2) to change the mains polarity.