48 -
|
Page Rigel Medical
62353+
user Manual V1.4
Press the
button (F1) to obtain the schematic representation of this test:
Warning
– never leave the earth bond, or any other probe attached to any
moving parts.
Warning
– never touch the appliance under test whilst testing is in
progress. Doing so could result in injury due to electric shock and / or
sudden movement of any mechanical parts on the appliance.
Refer to Page 4 for notes on precautions required for compliance with the
relevant safety standards.
Do not exceed the maximum permitted voltage of 30 V AC/DC
with respect to earth potential! Electric Shock danger!
Press
button (F3) to set the Test Duration, and RMS Pass/Fail limit. Use the
arrow keys to highlight and set. Press the
button (F4) when complete. Pressing
the
button (F3) will exit Settings mode without saving changes.
Press the
button (F2) to change the mains polarity.
For both Class I and Class II appliances, connect ALL Patient Connections or Applied
Parts to the Applied Part Module and fit the Appliance mains plug into the EUT outlet.