PART V TROUBLESHOOTING
5-22
ENGLISH
5.8 SPARK BETWEEN HIGH VOLTAGE TERMINAL AND SCREW
A spark occurs between the unit high voltage contact terminal and
screw around it.
Most of time, this problem is due to the moist air, no cell in the unit, or
the high voltage power contact terminal is too close to the screws on the
power contact board. If the high voltage is adjusted too high from the
power supply the spark may occur (Fig. 5.8-1).
To fix this problem, proceed with the following steps.
-Unplug the power cord.
-Open the bottom plate.
-Paste a insulation tape or clear tape beneath the unit high voltage
contact terminal or use hot glue gun to bond the screws (Fig. 5.8-2).
Fig.5.8-1 Sparks between the high voltage
contact terminal and the screw on the
power contact board.
Fig.5.8-2 Paste insulation or clear tape beneath the high
voltage contact power terminal or bond it with hot melt
adhesive.
Spark
Required tools
1-Phillips screwdriver
2-Hot glue gun