Remote Commands
R&S
®
ZNH
223
User Manual 1334.5985.02 ─ 02
Parameters:
<Detector>
POSitive | NEGative | SAMPle | RMS | AVERage | APEak
*RST:
APE
For more information, see chapter "Detectors".
Example:
DET POS
Sets the primary detector to "positive peak".
[SENSe:]DETector<1...2>[:FUNCtion]:AUTO
<State>
This command couples and decouples the detector to the trace mode.
The numeric suffix at DETector specifies the primary or secondary detector.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
DET:AUTO OFF
Turns off automatic detector selection for the primary detector.
12.6.2.6
Using Markers and Deltamarkers
Markers and Deltamarkers
The following commands control the operation of marker, delta markers and marker
functions.
List of commands
●
CALCulate<1...2>:DELTamarker<1...6>[:STATe]
●
CALCulate<1...2>:DELTamarker<1...6>:AOFF
●
CALCulate<1...2>:DELTamarker<1...6>:MAXimum[:PEAK]
●
CALCulate<1...2>:DELTamarker<1...6>:MAXimum:NEXT
●
CALCulate<1...2>:DELTamarker<1...6>:MINimum[:PEAK]
●
CALCulate<1...2>:DELTamarker<1...6>:X
●
CALCulate<1...2>:DELTamarker<1...6>:X:RELative
●
CALCulate<1...2>:DELTamarker<1...6>:Y?
●
CALCulate<1...2>:MARKer<1...6>[:STATe]
●
CALCulate<1...2>:MARKer<1...6>:AOFF
●
CALCulate<1...2>:MARKer<1...6>:MAXimum[:PEAK]
●
CALCulate<1...2>:MARKer<1...6>:MAXimum:NEXT
●
CALCulate<1...2>:MARKer<1...6>:MINimum[:PEAK]
●
CALCulate<1...2>:MARKer<1...6>:MODE
●
CALCulate<1...2>:MARKer<1...6>:X
Remote Control - Commands