Remote Commands
R&S
®
ZNH
264
User Manual 1334.5985.02 ─ 02
Suffix:
<1...2>
.
1...2
Parameters:
<YShiftFactor>
Numeric value in the range from -50 % to 50 %.
*RST:
0
Example:
DISP:ZOOM:Y -25
Shifts the zoom window 25% down.
12.6.3.7
Configuring and Using Measurement Functions
This chapter provides information on how to configure two-port measurements with the
tracking generator. The structure follows the order of the actual operation sequence
used when performing a measurement:
To perform the actual measurement, use the commands described in
Selecting the Measurement Port
...........................................................................264
Selecting the Cable Characteristics
...................................................................... 264
Selecting the Measurement Mode
........................................................................ 265
Selecting and Reading the Measurement
.............................................................266
................................................................................270
........................................................................................ 273
Selecting the Measurement Port
The following commands select the measurement port.
List of commands
........................................................................................... 264
MEASurement<1...2>:PORT
<PortNumber>
This command selects the measurement port.
The numeric suffix at MEASurement is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<PortNumber>
Numeric value in the range from 1 to 2.
*RST:
1
Example:
MEAS:PORT 1
Selects port 1.
Selecting the Cable Characteristics
The following commands define the cable characteristics for the cable you are testing.
Remote Control - Commands