Remote Commands
R&S
®
ZNH
217
User Manual 1334.5985.02 ─ 02
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value in the range from 1 dB to 1000 mrho.
The number you enter is rounded up to the next possible display
range. For example, if you enter 18, the R&S
ZNH automatically
sets the display range to 20 mrho.
Default unit: RHO
Example:
DISP:REFL:Y:SCAL 100
Sets a display range of 100 mrho.
INPut:ATTenuation<1...2>
<Attenuation>
This command defines the input attenuation.
The attenuation is coupled to the reference level. If you set the attenuation independ-
ently, the R&S
ZNH turns off this coupling.
The R&S
ZNH adjusts the reference level if it cannot be set for the current RF attenua-
tion.
Suffix:
<1...2>
.
1...2
RF port selection. Default selection at 1.
Parameters:
<Attenuation>
Range:
0 dB to 40 dB
*RST:
0 dB (AUTO is ON)
Default unit: dB
Example:
INP:ATT1 30dB
Defines RF attenuation of 30 dB at RF port 1 and deactivates
coupling to the reference level.
UNIT<1...2>:POWer
<Unit>
This command selects the unit of the vertical axis.
The availability of units depends on the operating mode and type of measurement.
The numeric suffix at UNIT is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<Unit>
DBM | DBMV | DBUV | DUVE | DMVE | VEMF | DBUA | VOLT |
WATT | DUVM | DUAM | V | W | V_M | W_M2 | DB | RHO |
MRHO
Note that the availability of units depends on the operating
mode.
*RST:
DBM
Remote Control - Commands