Remote Commands
R&S
®
ZNH
239
User Manual 1334.5985.02 ─ 02
If you set a span of 0 Hz, the R&S
ZNH starts a measurement in the time domain.
Parameters:
<FrequencySpan>
Range:
Specified in the data sheet.
*RST:
fmax with fmax = maximum frequency
Default unit: GHz
Example:
FREQ:SPAN 10MHz
Defines a span of 10 MHz.
[SENSe:]FREQuency:SPAN:AUTO
<State>
This command turns on or off the automatic calculation of the ideal span.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
FREQ:SPAN:AUTO ON
Turns on the automatic span determination.
[SENSe:]FREQuency:SPAN:FULL
This command restores the full span.
This command is an event and therefore has no query and no *RST value.
Example:
FREQ:SPAN:FULL
Restores full span.
Usage:
Event
[SENSe:]FREQuency:STARt
<StartFrequency>
This command defines the start frequency for measurements in the frequency domain
(span > 0).
Parameters:
<StartFrequency>
Range:
Depends on the operating mode and is specified in
the datasheet.
*RST:
0 Hz
Default unit: GHz
Example:
FREQ:STAR 20MHz
Defines a start frequency of 20 MHz.
[SENSe:]FREQuency:STOP
<StopFrequency>
This command defines the stop frequency for measurements in the frequency domain
(span > 0).
Remote Control - Commands