Working with the Measurement Wizard
R&S
®
ZNH
99
User Manual 1334.5985.02 ─ 02
Figure 6-3: Measurement specification dialog box
1 = Measurement step in a wizard set
2 = Individual measurement step status
Performing a sequence of measurements
Calibration data
The R&S
ZNH adds the calibration data of each calibration you perform to a calibration
data folder available on its internal memory. Each dataset in the calibration data folder
is unique for a particular calibration method and a particular measurement configura-
tion.
Note that the calibration is only valid for the instrument it has been performed on.
Before you start a measurement sequence with the measurement wizard, the
R&S
ZNH compares the contents of the calibration data folder to the measurement
configurations and required calibration methods in the measurement sequence.
●
If the R&S
ZNH has already been calibrated for a particular configuration, it
restores that data. Another calibration is not necessary.
●
If the R&S
ZNH has not been calibrated for a particular configuration yet, calibra-
tion becomes necessary. The new calibration data is added to the calibration data
folder.
Therefore, calibration is necessary only for calibration methods that have not been
used for a particular measurement configuration before.
However, it is recommended to perform calibration regularly to keep measurements as
accurate as possible.
You can delete obsolete calibration data from the folder:
\Public\Calibration Data
measurement setting parameters
concerning the mea-
surement task, you can start the measurement procedure.
Performing and Configuring Measurements