Remote Commands
R&S
®
ZNH
221
User Manual 1334.5985.02 ─ 02
Parameters:
<TriggerDelay>
Range:
0 s to 100 s
*RST:
0 s
Default unit: s
Example:
TRIG:HOLD 500us
Sets the trigger delay to 500 µs.
TRIGger[:SEQuence]:SLOPe
<TriggerSlope>
This command selects the slope of the trigger signal.
The trigger slope applies to all trigger sources.
Parameters:
<TriggerSlope>
POSitive | NEGative
*RST:
POSitive
Example:
TRIG:SLOP NEG
TRIGger[:SEQuence]:SOURce
<TriggerSource>
This command selects the trigger source.
Parameters:
<TriggerSource>
IMMediate | EXTernal | VIDeo | IQPower | GATed
IMMediate
Selects Free Run measurements.
EXTernal
Selects an external trigger.
VIDeo
Selects the video trigger.
IQPower
Selects an IQ Power trigger.
Gated
Selects a gated trigger.
*RST:
IMMediate
For more information, see chapter "Setting the Sweep".
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal.
12.6.2.5
Working with Traces
The following commands set up the trace and the various functions associated with it,
e.g. trace mathematics or the selection of the detector.
List of commands
●