Remote Commands
R&S
®
ZNH
289
User Manual 1334.5985.02 ─ 02
Parameters:
<FileName>
String containing the filename.
Example:
MMEM:NAME 'Public\Screeshots\Test.png'
Stores the screenshot in the corresponding directory on the
R&S
ZNH.
12.6.7
Configuring Data Capture
The following commands configure the data capture.
These commands are independent from the operating mode.
List of commands
........................................................................................... 289
SYSTem:CAPTure:DATaset[:STATe]
.................................................................................289
...............................................................................................289
................................................................................ 290
SYSTem:CAPTure:COUNter
<CaptureCounter>
This command defines the start of the filename counter.
The counter numbers the files stored when you capture data (screenshots, datasets
etc.).
Parameters:
<CaptureCounter>
String containing the number with which to start numbering files.
*RST:
'0000'
Example:
SYST:CAPT:COUN '0100'
Starts numbering files with 0100, e.g. Measurement0100.png.
SYSTem:CAPTure:DATaset[:STATe]
<State>
This command includes or excludes datasets from the data capture.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
SYST:CAPT:DAT ON
Includes datasets into the data capture.
SYSTem:CAPTure:MODE
<CaptureMode>
This command selects the data types that the R&S
ZNH saves when you capture the
current measurement data.
Parameters:
<CaptureMode>
SCReen | DATaset | BOTH