Remote Commands
R&S
®
ZNH
269
User Manual 1334.5985.02 ─ 02
Suffix:
<1...2>
.
1...2
Parameters:
<Return values>
Cable loss in dB.
Example:
CALC:TRAC:CABL:LOSS:RES?
Returns the cable loss in dB.
Usage:
Query only
CALCulate<1...2>:TRACe:LIMit:VSWR:FAIL?
This command queries the results of the limit check for the VSWR measurement for-
mat.
The numeric suffix at CALCulate is irrelevant for this command.
This command is a query and therefore has no *RST value.
Suffix:
<1...2>
.
1...2
Parameters:
<Return values>
0 | 1
0
Limit check fail.
1
Limit check pass.
Example:
CALC:TRAC:LIM:VSWR:FAIL?
Queries the results of the limit check.
Usage:
Query only
CALCulate<1...2>:TRACe:VSWR:AVG:RESult?
This command queries the average VSWR measurement result.
The command is available when the calculation of the VSWR has been turned on with
CALCulate<1...2>:TRACe:VSWR:MAX[:STATe]
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<Return values>
Average VSWR (no unit)
Example:
CALC:TRAC:VSWR:MAX:STAT ON
CALC:TRAC:VSWR:AVG:RES?
Queries the average VSWR.
Usage:
Query only
Remote Control - Commands