Remote Commands
R&S
®
ZNH
262
User Manual 1334.5985.02 ─ 02
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<MarkerPosition>
Numeric value of the marker position.
Example:
INIT:CONF OFF
CALC:MARK2 ON
Turns on single sweep mode and marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:Y?
Queries the position of marker 2.
Usage:
Query only
Marker Functions
The following commands perform various kinds of analysis at the marker position.
List of commands
DISPlay<1...2>:IMPedance:REFerence:POSition
.............................................................. 262
DISPlay<1...2>:ZOOM:AREA[:STATe]
.............................................................................. 262
........................................................................................ 263
.................................................................................................263
.................................................................................................263
DISPlay<1...2>:IMPedance:REFerence:POSition
<RefPosition>
This command sets the reference impedance for the smith chart measurement format.
The impedance can be between 1mΩ and 10 kΩ.
The numeric suffix at DISPlay selects the measurement screen in dual trace mode and
is in the range <1...2>.
Suffix:
<1...2>
.
1...2
Parameters:
<RefPosition>
Numeric value in the rnage from 1 mΩ to 10 kΩ.
*RST:
50
Default unit: OHM
Example:
DISP:IMP:REF:POS 75 OHM
Sets the reference impedance to 75 Ohm.
DISPlay<1...2>:ZOOM:AREA[:STATe]
<State>
This command turns the marker zoom function in a Smith chart on and off.
The numeric suffix at DISPlay selects the measurement screen in dual trace mode and
is in the range <1...2>.
Remote Control - Commands