Remote Commands
R&S
®
ZNH
257
User Manual 1334.5985.02 ─ 02
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Parameters:
<MarkerPosition>
Example:
INIT:CONT OFF
CALC:DELT2 ON
Turns on single sweep mode and delta marker 2.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2:Y?
Queries the position of delta marker 2.
Usage:
Query only
CALCulate<1...2>:MARKer<1...6>[:STATe]
<State>
This command turns markers on and off.
If you do not use a suffix at MARKer, marker 1 is selected. If one or more delta mark-
ers (2 to 6) are already active, the command turns these delta markers into normal
markers.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK3 ON
Turns on marker 3.
CALCulate<1...2>:MARKer<1...6>:AOFF
This command turns off all active markers, delta markers and active marker measure-
ment functions.
This command is an event and therefore has no query and no *RST value.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK:AOFF
Switches off all markers.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:FUNCtion:STRack[:STATe]
<State>
This command turns marker tracking on and off.
The numeric suffix at CALCulate and MARKer are irrelevant for this command.
Remote Control - Commands