Remote Control Commands
R&S
®
NRX
344
User Manual 1178.5566.02 ─ 07
Sensor parameter
Remote control command
Value
Trigger delay
TRIGger<Measurement>[:CHANnel<Channel>]:
DELay[:VALue]
0
s
Trigger level
TRIGger<Measurement>[:CHANnel<Channel>]:LEVel
-
30
dBm
Trigger source
TRIGger<Measurement>[:CHANnel<Channel>]:
SOURce
INT
Trigger slope
TRIGger<Measurement>[:CHANnel<Channel>]:SLOPe
POS
Trigger holdoff
TRIGger<Measurement>[:CHANnel<Channel>]:
HOLDoff
39.9
ms
Trigger dropout time
TRIGger<Measurement>[:CHANnel<Channel>]:DTIMe
0
s
Trigger hysteresis
TRIGger<Measurement>[:CHANnel<Channel>]:
HYSTeresis
3
dB
Table 13-8: WCDMA/3GPP FDD
Sensor parameter
Remote control command
Value
Sensor mode
CONTav
Aperture time
CALCulate<Measurement>[:CHANnel<Channel>][:
POWer][:AVG]:APERture[:VALue]
10
ms
Duty cycle correction
CALCulate<Measurement>[:CHANnel<Channel>]:
CORRection:DCYCle:STATe
OFF
Smoothing
CALCulate<Measurement>[:CHANnel<Channel>][:
POWer][:AVG]:SMOothing[:STATe]
OFF
Filter auto
CALCulate<Measurement>[:CHANnel<Channel>]:
AVERage:COUNt:AUTO[:STATe]
OFF
Filter length
CALCulate<Measurement>[:CHANnel<Channel>]:
AVERage:COUNt[:VALue]
16
Trigger source
TRIGger<Measurement>[:CHANnel<Channel>]:
SOURce
EXT
Table 13-9: WCDMA/3GPP TDD DL
Sensor parameter
Remote control command
Value
Sensor mode
TRACe
Aperture time
CALCulate<Measurement>[:CHANnel<Channel>][:
POWer][:AVG]:APERture[:VALue]
10
ms
Duty cycle
CALCulate<Measurement>[:CHANnel<Channel>]:
CORRection:DCYCle[:VALue]
6.667
%
Duty cycle correction
CALCulate<Measurement>[:CHANnel<Channel>]:
CORRection:DCYCle:STATe
ON
Smoothing
CALCulate<Measurement>[:CHANnel<Channel>][:
POWer][:AVG]:SMOothing[:STATe]
OFF
Configuring Sensors
www.allice.de
Allice Messtechnik GmbH