Measurement Basics
R&S
®
NRX
46
User Manual 1178.5566.02 ─ 07
a) Press the [Preset] key.
b) Tap "Preset".
See also
Chapter 9, "Saving and Recalling Settings"
2. Depending on the power sensor and the measurement conditions, consider to zero
the power sensor:
Execute zeroing:
Note:
Turn off all measurement signals before zeroing. An active measurement
signal during zeroing causes an error.
a) Switch off the power of the signal source.
b) Press the [Zero] key.
c) Tap "Zero All Sensors".
See also
3. Configure the measurement.
a) Open the "Measurement Settings" dialog, as described in
b) Select the "Measurement Type", for example "Continuous Average".
c) Tap "Quick Setup"
> "Auto Set".
4. Switch on the signal source.
The measurement starts, and the result is displayed in dBm.
5. If necessary, perform further settings.
5.4
Limit Violation
If a measured value violates the set limits, it is highlighted in red.
► To change the limit settings, tap the displayed limit values. See also (2) in
Limit Violation