CALCulate<Measurement>:LIMit<undef>:FCOunt?
Queries the number of limit violations. The counter is reset if one of the following
events occurs:
●
R&S NRX is switched on.
●
).
●
CALCulate<Measurement>:LIMit<undef>:CLEar[:IMMediate]
is execu-
ted.
Suffix:
<Measurement>
.
1 to 8
Measurement
<undef>
1 to n
No suffix required.
Usage:
Query only
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer[:DATA]
<value>
Sets the lower limit for the measured values.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = forward, 2 = reflection (reverse)
Parameters:
<value>
Manual operation:
See
See
"Forward Lower Limit, Reflection Lower Limit"
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer[:DATA]:CCDF
<value>
Sets the lower limit for the complementary cumulative distribution function (CCDF).
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
0.0
Default unit: pct
Manual operation:
See
"Forward Lower Limit, Reflection Lower Limit"