Sensor Configuration
R&S
®
NRX
120
User Manual 1178.5566.02 ─ 07
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:CORRection:
STATe
Gamma Phase
Sets the phase angle of the complex reflection coefficient of the source.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:PHASe
Gamma Magnitude
Sets the magnitude of the complex reflection coefficient of the source.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma[:MAGNitude]
8.3
Filter Settings
Access:
"Measurement Settings" > "Primary Sensor Config" > "Filter"
"Measurement Settings" > "Secondary Sensor Config" > "Filter"
Available for all measurement types.
Use the averaging filter to reduce fluctuations in the measurement results. Such fluctu-
ations can be caused by inherent noise of the power sensor, modulation of the mea-
surement signal or beats from the superposition of adjacent carriers. A more stable dis-
play is traded off against longer measurement times, caused by longer settling times
when the power changes. As a starting point, always use automatic filtering. If the
automatically selected filter setting proves to be not adequate, you can increase or
decrease the averaging number manually.
If you want to learn more about methods how to control the measurement, see the user
manual of the power sensor.