Remote Control Commands
R&S
®
NRX
326
User Manual 1178.5566.02 ─ 07
For some power sensors, the S-parameter correction is always activated. If you try to
deactivate the S-parameter correction for such a power sensor, a SCPI error is gener-
ated.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
ON | OFF
*RST:
OFF
Manual operation:
See
[SENSe<Sensor>:]RGAMma[:MAGNitude]
<magnitude>
Sets the magnitude of the complex reflection coefficient of the source ,
Γ
source
.
A value of 0.0 corresponds to an ideal matched source. A value of 1.0 corresponds to
total reflection.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<magnitude>
0.0
Disables the compensation.
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: -
[SENSe<Sensor>:]RGAMma:PHASe
<phase_angle>
Sets the phase angle of the complex reflection coefficient of the source,
Γ
source
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<phase_angle>
Range:
0.0 to 360.0
*RST:
0.0
Default unit: deg
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:STATe
<state>
Enables or disables the offset correction.
Configuring Sensors