Remote Control Commands
R&S
®
NRX
221
User Manual 1178.5566.02 ─ 07
TRIGger<undef>:ALL:COUNt
<count>
TRIGger<Measurement>[:CHANnel<Channel>]:COUNt
<count>
Sets the number of measurement cycles to be performed when the measurement is
started using
INITiate<Measurement>[:IMMediate]
.
This number equals the number of results that can be obtained from the sensor after a
single measurement. As long as the defined number of measurements is not executed,
the sensor automatically initiates another measurement internally when the current
result is available.
This command is particularly useful in conjunction with buffered measurements. For
example, to fill a buffer with a predefined size with measurements that have been trig-
gered externally or by
without having to start the measurement multiple times.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<count>
Range:
Depends on power sensor
*RST:
1
TRIGger<undef>:ALL:DELay:AUTO
<stat>
TRIGger<Measurement>[:CHANnel<Channel>]:DELay:AUTO
<stat>
If enabled, no measurement is started until the power sensor has settled. For this pur-
pose, the delay value is automatically determined.
If a longer period is set using
TRIGger<Measurement>[:CHANnel<Channel>]:
, the automatically determined delay is ignored.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<stat>
ON | OFF
*RST:
OFF
Measurement Settings and Results