Remote Control Commands
R&S
®
NRX
405
User Manual 1178.5566.02 ─ 07
SYSTem:SERRor:LIST:ALL?
Queries the list of all static errors that have occurred so far. The list is persistent. You
can remove entries using
Usage:
Query only
SYSTem:SERRor:LIST:NEXT?
Queries the list of all static errors that have occurred but have already been resolved
for the oldest entry and removes it from the queue. The response consists of an error
number and a short description of the error.
Usage:
Query only
SYSTem:SERRor:REMove
<num>
Removes an entry from the list of static errors. The entry is identified by its error num-
ber.
Setting parameters:
<num>
Unique sequence number
*RST:
0
Usage:
Setting only
13.13.10
Locking
SYSTem:KLOCk
<klock>
Disables or enables the front panel keyboard of the instrument including the [LOCAL]
key.
Parameters:
<klock>
OFF | ON
SYSTem:LOCK:SHARed:STRing?
Queries the lock string assigned to the locking group if shared locking is enabled.
Writes an error in the error queue if shared locking is not enabled.
Return values:
<result>
Usage:
Query only
System Information and Configuration
www.allice.de
Allice Messtechnik GmbH