Remote Control Commands
R&S
®
NRX
198
User Manual 1178.5566.02 ─ 07
Manual operation:
See
See
"Forward Scale Lower Limit, Reflection Scale Lower Limit"
CALCulate<Measurement>:METer<DirectionalChannel>:UPPer[:DATA]:CCDF
<value>
Sets the upper limit for the CCDF bargraph display.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
100.0
Default unit: pct
Manual operation:
See
"Forward Scale Upper Limit, Reflection Scale Upper Limit"
CALCulate<Measurement>:METer<DirectionalChannel>:UPPer[:DATA]:RATio:
RCOefficient
<value>
Sets the upper limit for the reflection coefficient bargraph display.
Suffix:
<Measurement>
.
1 to 8
Measurement
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
5.0
Default unit: -
Manual operation:
See
See
"Forward Scale Upper Limit, Reflection Scale Upper Limit"
CALCulate<Measurement>:METer<DirectionalChannel>:UPPer[:DATA]:RATio:
RFRatio
<value>
Sets the upper limit for the ratio of the forward/reverse power bargraph display.
Measurement Settings and Results