Remote Control Commands
R&S
®
NRX
289
User Manual 1178.5566.02 ─ 07
Parameters:
<expression>
See
UNIT<Measurement>:POWer[:VALue]
The query returns two additional expressions, see
*RST:
Depends on the selected channel.
Default unit: Depends on <expression> and the set unit.
Manual operation:
See
"Channel Calculation Function"
CALCulate<Measurement>:MATH[:EXPRession]:CATalog?
[<expressions>]
Queries all channel calculation functions supported by
Suffix:
<Measurement>
.
1 to 8
Measurement
Query parameters:
<expressions>
Usage:
Query only
Manual operation:
See
"Channel Calculation Function"
13.6.2
Using a Calculation Function
The following commands combine several setting commands and thus simplify pro-
gramming of the R&S NRX. They use parameter lists that differ for each measurement
type.
●
CONFigure
Configures according to the parameter list, but does not start a measurement.
The query without parameters, for example
CONF?
, returns the parameters trans-
fered the last time. Since the instrument settings can be changed after sensing a
CONFigure
command, the query does not return the current instrument setup.
●
READ
Compares the parameter list to the current settings, starts a measurement and
returns the result. If the parameter list does not match, a SCPI error is returned,
and the command is aborted.
●
MEASure
Configures according to the parameter list, starts a measurement and returns the
result. Thus, this command combines the
CONFigure
and
READ
commands.
The query without parameters, for example
MEAS?
, returns the parameters trans-
fered the last time.
●
FETCh
Returns the last valid measurement result.
Calculation Functions