Remote Control Commands
R&S
®
NRX
210
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>:LIMit<undef>:CLEar:AUTO
<mode>
If enabled, automatically resets the limit monitoring state and the internal counter for
limit violations if one of the following commands is executed:
●
INITiate<Measurement>[:IMMediate]
●
INITiate<Measurement>:CONTinuous
ON
●
MEASure<Measurement>...
query
●
READ<Measurement>...
query
Suffix:
<Measurement>
.
1 to 8
Measurement
<undef>
1 to n
No suffix required.
Parameters:
<mode>
OFF | ON | ONCE
CALCulate<Measurement>:LIMit<undef>:CLEar[:IMMediate]
Resets the limit monitoring state and the internal counter for limit violations.
Suffix:
<Measurement>
.
1 to 8
Measurement
<undef>
1 to n
No suffix required.
Usage:
Event
CALCulate<Measurement>:LIMit<undef>:FAIL?
Queries whether upper or lower limits have been violated. If one of the following events
occurs, the status is reset:
●
R&S NRX is switched on.
●
Reset is performed (
●
CALCulate<Measurement>:LIMit<undef>:CLEar[:IMMediate]
ted.
Suffix:
<Measurement>
.
1 to 8
Measurement
<undef>
1 to n
No suffix required.
Usage:
Query only
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH