Configuration for All Measurement Types
R&S
®
NRX
59
User Manual 1178.5566.02 ─ 05
Specific Trigger
Not available for each sensor type. Groups the specific trigger settings.
Jitter Suppression ← Specific Trigger
Defines the method how to cope with the misalignment between the trigger event and
the sample point.
"Compensate"
Compensation means resampling of trace result.
"Measure"
Does not perform resampling, but stores the measured trigger jitter.
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:JITTer:METHod
Trigger Master State ← Specific Trigger
Enables or disables the trigger master mode of the sensor. If enabled, the power sen-
sor outputs a digital trigger signal in sync with its own trigger event. The trigger signal
is output at the port selected under
The trigger master has to use its internal trigger source. Set the trigger source for the
trigger slaves to "Internal [A to D]", where [A to D] is the port to which the trigger mas-
ter is connected. The trigger signal generated by the trigger master is routed to the
R&S NRX and from there it is distributed to the trigger slaves and, if
is set to "Sensor [A to D]", to the trigger output.
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer[:STATe]
Trigger Master Port ← Specific Trigger
Sets the port where the trigger master sensor outputs a digital trigger signal.
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer:PORT
Trigger Synchronize State ← Specific Trigger
Usually used if "On" is set under
If enabled, blocks the external trigger bus as long as the sensor remains in the mea-
surement state. Thus, ensures that a new measurement is only started after all sen-
sors have completed their measurements.
Make sure that the number of repetitions is the same for all sensors involved in the
measurement. Otherwise, the trigger bus is blocked by any sensor that has completed
its measurements before the others and has returned to the idle state.
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:SYNChronize[:STATe]
Trigger Synchronize Port ← Specific Trigger
Sets the internal or external connection for the sync output of the sensor. For more
information, see
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:SYNChronize:PORT
Triggering