Remote Control Commands
R&S
®
NRX
193
User Manual 1178.5566.02 ─ 05
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<state>
*RST:
OFF
Manual operation:
See
See
"Forward Lower Limit State, Reflection Lower Limit State"
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer[:DATA]
<value>
Sets a lower limit for the measured values.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<DirectionalChannel> 1 to 2
1 = forward, 2 = reflection (reverse)
Parameters:
<value>
Manual operation:
See
"Forward Lower Limit, Reflection Lower Limit"
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer[:DATA]:CCDF
<value>
Lower limit value for CCDF.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
0.0 to 100.0
*RST:
0.0
Default unit: pct
Manual operation:
See
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer[:DATA]:RATio:
RCOefficient
<value>
Lower limit value for reflection coefficient.
Measurement Settings and Results