Sensor Configuration
R&S
®
NRX
113
User Manual 1178.5566.02 ─ 05
Filter Length
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
If
is set to "Auto", this parameter is read-only.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:ENUM
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:COUNt[:
VALue]
Recalc Filter Length
Available for continuous average, burst average, time gate measurements.
Recalculates the number of readings that are averaged for one measured value.
Clear Filter Buffer
Available for continuous average, burst average, time gate measurements.
Clears the filter buffer.
Remote command:
Fixed Noise Mode
Available for continuous average, burst average, time gate measurements.
Sets the autofilter.
"Normal"
Sets the averaging number so that the intrinsic noise of the power
sensor, 2 standard deviations, does not exceed the specified
"Fixed Noise"
Limits the averaging number as specified in
to avoid very long settling times.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
TYPE
Noise Content
If
is set to "Normal", available for continuous average, burst aver-
age, time gate measurements.
Sets the averaging number so that the intrinsic noise of the power sensor does not
exceed the specified value.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
NSRatio
Maximum Settling Time
If
is set to "Fixed Noise", available for continuous average, burst
average, time gate measurements.
Sets an upper time limit, a maximum time, that is never exceeded.
Filter Settings