Remote Control Commands
R&S
®
NRX
278
User Manual 1178.5566.02 ─ 05
Query parameters:
<expected_value_or_source_list>
<resolution_or_source_list>
<source_list>
<expr>
Usage:
Query only
CONFigure<Measurement>[:SCALar][:POWer][:AVG]:RLOSs?
[<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
FETCh<Measurement>[:SCALar][:POWer][:AVG]:RLOSs?
[<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
READ<Measurement>[:SCALar][:POWer][:AVG]:RLOSs?
[<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
MEASure<Measurement>[:SCALar][:POWer][:AVG]:RLOSs?
[<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
Used for
CALCulate<Measurement>:MATH[:EXPRession]
"RLOS(SENSn,SENSm)"
.
Return loss/transmission loss of a DUT, measured by two power sensors. The used
parameters are described in
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Query parameters:
<expected_value_or_source_list>
<resolution_or_source_list>
<source_list>
<expr>
Usage:
Query only
14.5.2.2
Continuous Average Calculation Functions with Buffering
Parameter list
The following parameters are used.
●
<buffered_size>
Mandatory. Number of requested measured values.
Corresponds to
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
.
●
<expected_value>
Optional. Value that is expected for the measurement.
●
<resolution>
Optional. Limit up to which the measurement result should be free of noise.
Corresponds to
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
.
●
<source_list>
Calculation Functions