Remote Control Commands
R&S
®
NRX
370
User Manual 1178.5566.02 ─ 05
SYSTem:ERRor:COUNt?
Queries the number of entries in the error queue.
Return values:
<ErrorCount>
Example:
SYSTem:ERRor:COUNt
Queries the number of entries in the error queue.
Response: 1
One error has occurred since the error queue was last read out.
Usage:
Query only
SYSTem:ERRor:EXTended[:STATe]
<state>
Enables or disables the extended error messages, with more information than
. The location of errors in the parsed command strings is
also shown.
Parameters:
<state>
*RST:
ON
SYSTem:ERRor[:NEXT]?
Queries the error/event queue for the oldest item and removes it from the queue. The
response consists of an error number and a short error description.
Return values:
<ErrorCode>
Positive error numbers are instrument-dependent. Negative
error numbers are reserved by the SCPI standard.
0
No errors have occurred since the error queue was last read out.
<ErrorDescription>
Short error description
Usage:
Query only
SYSTem:SERRor:LIST:ALL?
Queries the list of all static errors that have occurred so far. The list is persistent. You
can remove entries using
Usage:
Query only
SYSTem:SERRor:LIST:NEXT?
Queries the list of all static errors that have occurred but have already been resolved
for the oldest entry and removes it from the queue. The response consists of an error
number and a short description of the error.
Usage:
Query only
System Information and Configuration