Remote Control Commands
R&S
®
NRX
233
User Manual 1178.5566.02 ─ 05
[SENSe<Sensor>:]TRACe:MEASurement:OFFSet:TIME
<value>
Sets the length of the gate.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
Default unit: s
[SENSe<Sensor>:]TRACe:MID:OFFSet:TIME
<time>
Sets the length of the gate in which the pulse analysis is performed.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
0.0 to 30.0
*RST:
0.01
Default unit: s
[SENSe<Sensor>:]TRACe:MID:TIME
<time>
Sets the start time of the gate in which the pulse analysis is performed.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
0.0 to 30.0
*RST:
0.01
Default unit: s
CALCulate<Measurement>:TRACe:MEASurement:ALGorithm
<value>
Effective for pulse analysis measurements.
Sets the analysis algorithm for detecting the pulse top and the pulse base power of a
pulsed signal. From these two power levels, the reference levels are derived.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Measurement Settings and Results