Remote Control Commands
R&S
®
NRX
403
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>:STATistics[:SCALe]:X:POINts
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<points>
Range:
3 to 8191
*RST:
200
[SENSe<Sensor>:]STATistics:SCALe:X:RANGe
<range>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>:STATistics[:SCALe]:X:RANGe
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<range>
Range:
0.01 to 100.0
*RST:
50.0
Default unit: dB
[SENSe<Sensor>:]STATistics:SCALe:X:RLEVel
<rlev>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>:STATistics[:SCALe]:X:RLEVel:RELative
CALCulate<Measurement>:STATistics[:SCALe]:X:RLEVel[:ABSolute]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<rlev>
Range:
-280.0 to 220.0
*RST:
-30.0
Default unit: dBm
[SENSe<Sensor>:]STATistics:TIME
<time>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:TIME
if a gate is set.
CALCulate<Measurement>:STATistics:APERture
if no gate is set.
R&S NRP2 Compatibility