CALCulate<Measurement>:TSLot:TIMing:EXCLude:STARt
................................................248
CALCulate<Measurement>:TSLot:TIMing:EXCLude:STOP
................................................ 248
CALCulate<Measurement>[:POWer]:TSLot[:AVG]:COUNt
..................................................248
CALCulate<Measurement>[:POWer]:TSLot[:AVG]:SELection
............................................. 249
CALCulate<Measurement>[:POWer]:TSLot[:AVG]:WIDTh
..................................................249
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID:OFFSet[:TIME]
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID:TIME
.............................249
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID[:STATe]
CALCulate<Measurement>:TSLot:TIMing:EXCLude:STARt
<value>
Sets the time that is excluded at the beginning of the integration period.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
Range:
0.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>:TSLot:TIMing:EXCLude:STOP
<value>
Sets the time that is excluded at the end of the integration period.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
Range:
0.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TSLot[:AVG]:COUNt
<value>
Sets the number of simultaneously measured timeslots.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
Range:
1 to 128
*RST:
8
Manual operation:
See
Measurement Settings and Results