Remote Control Commands
R&S
®
NRX
227
User Manual 1178.5566.02 ─ 05
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MARKer<Marker>:MODE
<mode>
Sets the mode for the selected marker.
Suffix:
<Window>
.
1 to 4
Measurement channel
<Marker>
1 to 4
Marker (M1 to M4)
Parameters:
<mode>
OFF | RULer | MEASure
OFF
The measurement is off.
RULer
Draws a line ate the power or time position of the marker.
MEASure
Measures power (ratio) or time (difference).
*RST:
OFF
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MARKer<Marker>:POSition:MODE
<mode>
Defines where the marker is placed.
Suffix:
<Window>
.
1 to 4
Measurement channel
<Marker>
1 to 4
Marker (M1 to M4)
Parameters:
<mode>
FTIMe | FPOWer | RPOSition | RPOWer | RPLeft | RPRight |
PSEarch | MSEarch | RPSLeft | RPSRight | RMSLeft |
RMSRight
FTIMe
Fixed time
FPOWer
Fixed power
RPOSition
Relative to reference position
RPOWer
Relative to reference power
RPLeft
From reference power left
Measurement Settings and Results