Remote Control Commands
R&S
®
NRX
293
User Manual 1178.5566.02 ─ 05
Query parameters:
<dtolerance>
Default unit: s
<start_exclude>
Default unit: s
<end_exclude>
Default unit: s
<expected_value_or_source_list>
<resolution_or_source_list>
<source_list>
<expr>
Usage:
Query only
14.5.2.4
Timeslot Calculation Functions
Parameter list
For the calculation functions of the timeslot measurement, the following parameters are
used.
●
<tslot_width>
Mandatory. Width of a timeslot.
Corresponds to
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh
●
<no_slots>
Mandatory. Number of timeslots to be measured.
Corresponds to
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:COUNt
●
<start_exclude>
Mandatory. Amount of time at the beginning of a timeslot or integration period that
is not evaluated.
Corresponds to
[SENSe<Sensor>:]TIMing:EXCLude:STARt
.
●
<end_exclude>
Mandatory. Amount of time at the end of a timeslot or integration period that is not
evaluated.
Corresponds to
[SENSe<Sensor>:]TIMing:EXCLude:STOP
.
●
<expected_value>
Optional. Value that is expected for the measurement.
●
<resolution>
Optional. Limit up to which the measurement result should be free of noise.
Corresponds to
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
.
●
<source_list>
Optional. Primary and secondary sensor. The number for each sensor is preceded
by the character
@
. The entire expression is enclosed in parentheses.
Example: (
@
3),(
@
2)
Sensor C is the primary sensor, and sensor
B is the secondary sensor.
Calculation Functions