Remote Control Commands
R&S
®
NRX
307
User Manual 1178.5566.02 ─ 05
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
0.0 to 0.3
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:APERture[:
VALue]
<value>
Effective for continuous average measurements.
Sets the duration of the sampling window. During this time interval, the average signal
power is measured.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
8.3e-9 to 30.0
*RST:
0.01
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:SMOothing[:
STATe]
<value>
Enables or disables digital lowpass filtering of the sampled video signal.
The problem of instable display values due to a modulation of a test signal described
under
can also be eliminated by lowpass filtering of
the video signal. The lowpass filter eliminates the variations of the display even in case
of unperiodic modulation and does not require any other setting.
If modulation is periodic, setting the aperture time is the better method, since it allows
shorter measurement times.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Configuring Sensors