Remote Control Commands
R&S
®
NRX
314
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:MTIMe
<value>
Sets an upper time limit (maximum time) that is never exceeded. If
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
TYPE
NSR
is set, the R&S NRX has to determine the filter length automatically. The fil-
ter length can become large and thus also the measurement time. By setting an upper
time limit, you can prevent undesired long measurement times.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
0.1 to 1000.0
*RST:
1.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
NSRatio
<value>
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:
is set to
NSR
.
Sets the averaging number so that the intrinsic noise of the power sensor does not
exceed the specified value.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
100e-6 to 1.0
*RST:
1.0
Default unit: dB
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:SLOT
<value>
Sets a timeslot from which the measured value is used to determine the filter length
automatically. The timeslot number must not exceed the number of the currently set
timeslots.