Remote Control Commands
R&S
®
NRX
215
User Manual 1178.5566.02 ─ 05
TRIGger<undef>:ALL:SYNChronize[:STATe]
<state>
TRIGger<Measurement>[:CHANnel<Channel>]:SYNChronize[:STATe]
<state>
Usually used if
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer[:
is enabled.
If enabled, blocks the external trigger bus as long as the sensor remains in the mea-
surement state. Thus, ensures that a new measurement is only started after all sen-
sors have completed their measurements.
Make sure that the number of repetitions is the same for all sensors involved in the
measurement. Otherwise, the trigger bus is blocked by any sensor that has completed
its measurements before the others and has returned to the idle state.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
ON | OFF
*RST:
OFF
Manual operation:
See
TRIGger<undef>:ALL[:IMMediate]
TRIGger<Measurement>[:IMMediate]
This command triggers a measurement.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Usage:
Event
14.4.3
Selecting the Measurement
Before starting a measurement, select the measurement type.
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"<string>"
Determines the data that are processed. The parameters depend on the measurement
type.
Measurement Settings and Results