Remote Control Commands
R&S
®
NRX
212
User Manual 1178.5566.02 ─ 05
Parameters:
<level>
Range:
-290.0 to +223.01
*RST:
-10.0
Default unit: dBm
Manual operation:
See
TRIGger<undef>:ALL:MODE
<mode>
TRIGger<Measurement>:MODE
<mode>
Controls the trigger execution depending on the setting of the trigger source,
TRIGger<Measurement>[:CHANnel<Channel>]:SOURce
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<mode>
NORMal | FREerun | SINGle | AUTO
NORMal
Continuous triggering with regular trigger events.
FREerun
Automatically starts a measurement if
TRIGger<Measurement>[:CHANnel<Channel>]:SOURce
IMMediate
is set.
If another trigger source is set, this setting is automatically
changed to
AUTO
.
SINGle
Disables continuous triggering so that only one trigger event at a
time is executed.
AUTO
Repetitive, automatic trigger event.
*RST:
AUTO
Manual operation:
See
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer:PORT
<port>
Effective only if the connected sensor is trigger master, see
TRIGger<Measurement>[:CHANnel<Channel>]:MASTer[:STATe]
Sets the port where the trigger master sensor outputs a digital trigger signal.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Measurement Settings and Results